MD-9800E Mammography System

Products > High Frequency X-ray > MD-9800E Mammography System

MD-9800E Mammography System

The 9800 series high frequency Molybdenum target mammography machine are featured with small focus, high resolution, low dosage and micro computer controlled. It's the high performance breast screening machine for iconography examination of mammary diseases and with high sensitivity for mammary begin lump, nodes, calcified point and early detection of malignant
 
1: Easy to use, user interface design with finger touch operation panel, fast reachable button and foot switch
2: Streamline compressor plate provide the intact breast image
3: High quality carbon fiber cartridge table with low x ray absorption that reduces scattered x ray
 
 The 40KHz high frequency generator designed with advanced ICBT high frequency inversion technology and quasi resonant pressure boosting technology. It provides high efficiency, stable voltage, pure output spectral line that produces high definition image.
 
 Reliable and durable mechanical design provides:
1: Electrical up and down movement smoothly and quiet
2: 360 degree rotation C arm provides imaging flexibility
 
Specification:
 
X-ray Generator
Generator Type:High Frequency Inverter 40kHz
Input Power: Single phase 220VAC, 50/60Hz
Radiographic Ratings:Large Focal Point 20-35kV/10-510mAs
                     Small Focal Point 20-35kV/10-100mAs
Power Rating:3.6kW
 
X-ray Tube
Model:IAE XM15
Focal Spot Size: Dual Focus 0.1 / 0.3mm
Target Material:Molybdenum (Mo)
Port Material:Beryllium (Be)
Anode Heat Storage: 300kHU
Anode Cooling:Air cooling
Filtration: Mo(0.03mm),Al(0.5mm)
 
Radiographic Stand
C-ARM:Vertical Movement:580mm
Rotations Deg:+175°~-180°
SID: 650mm
 
Cassette Image Receptor
Bucky Device:18×24cm Bucky drive mechanism
Grid Ratio:5:1 ,30 Line/cm
 
Line Voltage
220Vac@25A,Single phase
 
Weight
240kg
 
Comfortable Compression
When some degree of pressure is required for radiography, it allows you to presser the appropriate pressure(up to a maximum of 20kg)and is equipped with MICOM Control’s Soft-touch system which is designed to minimize the discomfort of the examine with in the pressure range.
 
Tissue Compression: Manual and Motorized (Max 20kg)
                   Compression Force and Thickness Data Display
                     Micro Control’s Compression
                   Automatic Release
 
Intelligent Automatic Exposure ControlAEC
With the Automatic Exposure Control system, it is possible to produce images with reliable intensity suitable for and film, screen, or method of radiography.
 
Furthermore, it greatly enhances the convenience of radiography by embedding the Full-AEC function which is capable of utilizing the Auto kV
Type:Solid-State Detector or PMT Detector
      Microprocessor Control
AEC Mode:Full AEC(Auto kV)
Semi AEC (kV Select)
Manual (kV,mAs Select)
 

Density Adjustment:16 density steps